Part of Anesthesia Machine | |
Main body | High-strength engineering plastic rack, light, |
beautiful and corrosion resistant | |
Scope of application | adult, child |
Gas source | O2: 0.27~0.55MPa N2O: 0.27~0.55MPa |
Flow meter | O2: 0.1~1.0L/min 1.1~10L/min |
N20: | 0.1~1.0L/min 1.1~10L/min |
O2, N2Olinkage and N2O Stopper | |
When using nitrous oxide, oxygen concentration>25%; | |
When the oxygen pressure<0.2Kpa, the flow of nitrous oxide would be cut off. | |
Flow rate of rapid oxygen supply | 25~75L/min |
Low oxygen pressurealarm There will be sound alarm when the oxygen | |
pressure | < 0.2MPa |
Vaporizer it has the function of automatic compensation based on pressure, temperature, andflow rate.The regulation range of evaporator concentrationis 0~5 vol%. | |
Among Halothane, Enflurane, Isoflurane and Sevoflurane, two can be chosen for application as required by the customer. And vaporizers that are imported with original packaging are also available. | |
Respiratory circuit | working mode: all-close, semi-close, semi-open |
APL: | 0.5~7 kPa |
Respiratory Bellows bellows for adults, bellows for children | |
tidal volume range: | 0~1, 500 ml |
Part of Ventilator | |
Display mode | High-definition 5.7” LCD screen display |
Ventilation mode | IPPV, SIPPV, SIMV, PEEP, MANUAL, SIGH |
Ventilation parameters | Tidal volume 50~1500 ml |
Rate | 2~99 bpm |
SIMV rate | 2~20 bpm |
I:E | 2:1~1:8 |
Inspiratory trigger pressure | "-1.0~2.0 kPa" |
PEEP | |
Pressure range | 1.0~6.0 kPa |
SIGH | 1.5 times the inspiratory time |
Parameters for ventilation monitoring tidal volume, ventilation volume, IPPV rate, SIMV rate, total respiratory rate, I/E, peak pressure of airway, pressure - time waveform, flow rate - time waveform, PEEP, inspiratory trigger pressure | |
Monitoring of oxygen concentration | (21%~100%) |
Security Alarm System | |
Oxygen Concentration alarm | upper limit setting range (50%~100%) |
low limit setting range | (15%~50%) |
Airway pressure alarm | upper limit setting range 1.0~6.0 kPa |
low limit setting range | 0.4~2.0 kPa |
Per-minute ventilation volume alarm | upper limit setting range 3.0~30L/min |
low limit setting range | 1.0~10L/min |
Sustained high-pressure alarm | it will give alarm when stress have consistently |
been higher than 2.5 kPa | |
Suffocation alarm | it will give sound and light alarm if there is no tidal volume input for 15 seconds |
Power alarm | |
Power | AC 220V±10% 50Hz±1Hz |
(UPS, Storage battery) | |
(Mechanical Arms) | |
*Items within brackets are optional. |
Anesthesia Machine, with its ergonomic design, integrates a concise and smooth structure and high mobility. The microcomputer-controlled Ventilator adopts exported sensors and other components. Its scientific and proper design as well as rigorous production process are what demonstrate the inner quality of Aokai products: mechanical ventilation is made secure and visible by the folded respiratory bellows, and the security of air supply is guaranteed by pneumatic transmission and measurement system. With multiple modern technologies applied, the Anesthesia Machine is both outstanding in performance and versatile in functions.